Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
CCS120 - Analog total chlorine sensor for water and wastewater applications

Analog total chlorine sensor
CCS120

Disinfection sensor for all water and wastewater applications

CCS120
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Total chlorine

  • Application

    Potable water,Process water, Waste water, pools

  • Characteristic

    Amperometric measurement of total chlorine.

  • Measurement range

    0,1 - 10ppm total chlorine; pH 6,5 - 9,5

  • Measuring principle

    - Total chlorine consists of free chlorine (HOCl, OCl) and bounded chlorine (chloramines).
    - A constant polarisation voltage between working and counter electrode instigates the electrochemical reaction.
    - Only slightly pH dependent.

  • Design

    Closed amperometric two electrode sensor for continuous measurement of total chlorine.

  • Material

    Sensor shaft : PVC
    Membrane cap : PPE

  • Dimension

    Diameter : appr. 25mm
    Length : 210mm

  • Process temperature

    5°C - 45°C

  • Process pressure

    0bar - 3bar

  • Temperature sensor

    Including NTC-temperature sensor.

  • Connection

    TOP68 plug-in head