Specs at a glance
-
Accuracy
Standard: 0.05%
Platinum: up to 0.025% -
Process temperature
-40°C...125°C
(-40°F...257°F) -
Pressure measuring range
400 mbar...700 bar
(5.8 psi...10.500 psi) -
Main wetted parts
Alloy C276
316L
Rhodium -
Material process membrane
316L, AlloyC,
Rhodium> Gold -
Measuring cell
400 mbar...700 bar
(5.8 psi...10.500 psi)
relative/ absolute
Field of application
The Cerabar PMP71 digital pressure transmitter with metallic membrane is typically used in process applications for pressure, level, volume or mass measurement in liquids or gases. PMP71 is designed for high pressure applications up to 700bar. Quick Setup with freely adjustable measuring range without pressure specification. Designed according to IEC 61508 for use in SIL2/3 safety systems and available with MID Parts Certificate suitable for custody transfer applications.
Two chamber aluminum or stainless steel (316L) housing.
Process connections: Threads and flange connections
Process temperature: -40 to +125°C (-40 to +257°F)
Measuring ranges: -1/0 up to 700bar (-15/0 up to +10,500psi)
Accuracy: ±0.05%, "Platinum" ±0.025%
International explosion protection certificates, overfill prevention WHG, hygienic approvals, marine approvals, SIL
Benefits
Best accuracy, reproducibility and long-term stability
Highest safety due to gas tight feedthrough with capabilities up to SIL2/3, certified to IEC 61508
Easy menu-guided commissioning via local display, 4 to 20mA with HART, PROFIBUS PA, FOUNDATION Fieldbus
HistoROM data management concept for fast and easy commissioning, maintenance and diagnostics
Overload-resistant and function-monitored from the measuring cell to the electronics
Available with mounted manifolds: always fit, always tested for leaks
Seamless and independent system integration (HART/PA/FF)
Videos
-
Pressure measuring principle
Extended selection
-
High-end products
- Highly functional and convenient
Technical excellence
Simplicity

-
Standard products
- Reliable, robust and low-maintenance
Technical excellence
Simplicity
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable
FLEX selections | Technical excellence | Simplicity |
---|---|---|
Fundamental selection Meet your basic measurement needs |
Technical excellence
|
Simplicity
|
Lean selection Handle your core processes easily |
Technical excellence
|
Simplicity
|
Extended selection Optimize your processes with innovative technologies |
Technical excellence
|
Simplicity
|
Xpert selection Master your most challenging applications |
Technical excellence
|
Simplicity
Variable |
-
Measuring principle
Absolute and gauge pressure
-
Characteristic
Digital transmitter with piezoresistive sensor and welded metallic membrane
Modular transmitter
Long term stability
Enhanced safety via self diagnostic functionsSecondary process barrier -
Supply voltage
4...20 mA HART
10,5...45V DC (Non Ex):
Ex ia: 10,5...30V DC
PROFIBUS PA:
9...32 V DC (Non Ex)
FOUNDATION Fieldbus:
9...32 V DC (Non Ex) -
Reference Accuracy
Standard: 0.05%
Platinum: up to 0.025% -
Long term stability
0.05 % of URL/ year
0.07 % of URL/ 5 years
0.1 % of URL/ 10 years -
Process temperature
-40°C...125°C
(-40°F...257°F) -
Ambient temperature
-50°C...85°C
(-58°F...185°F) -
Measuring cell
400 mbar...700 bar
(5.8 psi...10.500 psi)
relative/ absolute -
Smallest calibratable span
5 mbar (0.075 psi)
-
Vacuum resistance
10 mbar (0.15 psi)
-
Max. Turn down
100:1
-
Max. overpressure limit
1050 bar (15.750 psi)
-
Process connection
Thread:
G1/2...G2, R1/2, MNPT1/2...MNPT2, R1/2
Flange:
DN25...DN80,
ASME 1"...4",
JIS 10K...20K -
Material process membrane
316L, AlloyC,
Rhodium> Gold -
Material gasket
None, measuring cell welded
-
Fill fluid
Silicone oil,
Inert oil -
Material housing
Die-cast aluminum,
AISI 316L -
Communication
4...20 mA HART
1- 5V DC Low Power
PROFIBUS PA
FOUNDATION Fieldbus -
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC, UK Ex
-
Safety approvals
SIL
-
Design approvals
EN10204-3.1
NACE MR0175 -
Marine approvals
GL/ABS
-
Drinking water approvals
NSF
-
Specialities
Diagnostic functions
-
Successor
PMP71B
-
Measuring principle
Absolute and gauge pressure
-
Characteristic / Application
Digital transmitter with piezoresistive sensor and welded metallic membrane
Modular transmitter
Long term stability
Enhanced safety via self diagnostic functions
Secondary process barrier -
Specialities
diagnostic functionalities
different languages in software -
Supply / Communication
4...20mA HART:
10,5...45V DC
Ex ia: 10,5...30V DC
PROFIBUS PA /
FOUNDATION Fieldbus:
9...32V DC -
Accuracy
Standard: 0.05%
Platinum: up to 0.025% -
Long term stability
0,05% of URL/year
-
Ambient temperature
-50°C...85°C
(-58°F...185°F) -
Process temperature
-40°C...125°C
(-40°F...257°F) -
Process pressure / max. overpressure limit
1050bar (15,750psi)
-
Pressure measuring range
400 mbar...700 bar
(5.8 psi...10.500 psi) -
Main wetted parts
Alloy C276
316L
Rhodium -
Process connection
Threads
Flanges (DIN, ASME, ...) with flush membrane -
Max. measurement distance
7000m (22.966ft) H2O
-
Communication
4...20 mA HART
1- 5V DC Low Power
PROFIBUS PA
FOUNDATION Fieldbus -
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC
-
Safety approvals
SIL
-
Design approvals
EN 10204-3.1
NACE MR0175, MR0103 -
Marine approval
GL/ ABS
-
Drinking water approvals
NSF
-
Options
HistoROM/M-Dat
4-line digital display
SS- or Aluminiumhousing
Separate housing -
Successor
PMP71B
-
Application limits
Measuring cell: Metal
welded
If pressurized, possibly
use differential pressure
meas-urement with two
pressure transmitters
(electronic dp). Observe
ratio head pressure :
hydrostatic pressure
-
-
Technical Information (TI)
TI Cerabar S PMC71, PMP71, PMP75
English version - 04/2024
New version available in English
Process pressure measurement 1-5 V DC, HART, PA, FF
EN 01/04/202413/01/202229/09/202001/04/202016/04/2019Language:
English
Version:
01/04/2024
File size:
3.5 MB
File name:
TI00383PEN_3624-00.pdf
-
-
-
Operating Instruction (BA)
Cerabar S PMC71, PMP71, PMP75 HART V02.30.zz Operating Instructions
English version - 08/2024
New version available in English
Pressure transmitter with ceramic and metal sensors
EN 01/08/202401/03/201810/12/201401/11/201407/05/2014Language:
English
Version:
01/08/2024
File size:
914.5 KB
File name:
BA00271PEN_2022.pdf
-
-
-
Technical Information (TI)
TI Weld-in adapter, process adapter and flanges
-
Technical Information (TI)
English version - 12/2024Level, pressure and temperature measurement
EN 06/12/202425/03/202415/09/202328/10/202130/07/2021Language:
English
Version:
06/12/2024
File size:
1.3 MB
File name:
TI00426FEN_2624.pdf
-
-
-
Technical Information (TI)
TI Cerabar S PMC71, PMP71, PMP75
-
Technical Information (TI)
English version - 04/2024Process pressure measurement
1-5 V DC, HART, PA, FFEN 01/04/202413/01/202229/09/202001/04/202016/04/2019Language:
English
Version:
01/04/2024
File size:
3.5 MB
File name:
TI00383PEN_3624-00.pdf
-
-
-
Technical Information (TI)
Cerabar S PMP71 with MID Part Certificate
-
Technical Information (TI)
English version - 12/2011Pressure transmitter with metal sensors; Overload-resistant and
function-monitored; Communication via HART; Subcomponent
suitablefEN 16/12/2011Language:
English
Version:
16/12/2011
File size:
1.5 MB
File name:
TI00451PEN_0111.pdf
-
-
-
Operating Instruction (BA)
Cerabar S PMP71 Low Power 1-5V DC Output Operating Instructions
-
Operating Instruction (BA)
English version - 08/2017Low power pressure transmitter with metal sensors and 1-5V DC voltage
outputEN 25/08/201701/08/2017Language:
English
Version:
25/08/2017
File size:
1.1 MB
File name:
BA01633PEN_0317.pdf
-
-
-
Operating Instruction (BA)
Description of Device Functions HART (V02.10.54) Cerabar S PM71 with MID Part Certificate
-
Operating Instruction (BA)
English version - 04/2017Pressure transmitter with metal sensors; Overload-resistant and
function-monitored; Communication via HART; Subcomponent
suitablefor custody transfer measurement with NMi approvalEN 03/04/201710/12/201424/04/201219/12/2011Language:
English
Version:
03/04/2017
File size:
1.3 MB
File name:
BA00413PEN_1316.pdf
-
-
-
Operating Instruction (BA)
Cerabar S PMC71, PMP71, PMP75 HART V02.30.zz Operating Instructions
-
Operating Instruction (BA)
English version - 08/2024Pressure transmitter with ceramic and metal sensors
EN 01/08/202401/03/201810/12/201401/11/201407/05/2014Language:
English
Version:
01/08/2024
File size:
914.5 KB
File name:
BA00271PEN_2022.pdf
-
-
-
Operating Instruction (BA)
Cerabar S PMC71, PMP71, PMP75 FOUNDATION Fieldbus (V 04.00.zz) Operating Instructions
-
Operating Instruction (BA)
English version - 08/2024Pressure transmitter with ceramic and metal sensors
Overload-resistant and function-monitored;
Communication via
FOUNDA
TION FieldbusEN 01/08/202403/04/201710/12/201407/05/201430/04/2013Language:
English
Version:
01/08/2024
File size:
1.9 MB
File name:
BA00302PEN_1922-00.pdf
-
-
-
Operating Instruction (BA)
Cerabar S / Deltabar S / Deltapilot S, Description of device functions FOUNDATION Fieldbus V 04.00.zz
-
Operating Instruction (BA)
English version - 04/2017Cerabar S PMC71, PMP71, PMP75 Pressure transmitter
Deltabar S FMD77, FMD78, PMD75 Differential pressure
transmitter
Deltapilot S FMB70 Hydrostatic pressure transmitterEN 03/04/201710/12/201407/05/201430/04/201301/10/2011Language:
English
Version:
03/04/2017
File size:
6.3 MB
File name:
BA00303PEN_1716.pdf
-
-
-
Operating Instruction (BA)
Cerabar S /Deltabar S /Deltapilot S Description of device functions PROFIBUS PA V04.01.zz
-
Operating Instruction (BA)
English version - 04/2017Cerabar S PMC71, PMP71, PMP75 Pressure transmitter
Deltabar S FMD77, FMD78, PMD75 Differential pressure
transmitter
Deltapilot S FMB70 Hydrostatic pressure transmitterEN 03/04/201710/12/201407/08/201301/10/201101/05/2010Language:
English
Version:
03/04/2017
File size:
4.8 MB
File name:
BA00296PEN_1616.pdf
-
-
-
Operating Instruction (BA)
Cerabar S PMC71, PMP71, PMP75 PROFIBUS PA (V 04.01.zz) Operating Instructions
-
Operating Instruction (BA)
English version - 08/2024Pressure transmitter with ceramic and metal sensors
Overload-resistant and function-monitored;
Communication via
PROFIB
USPAEN 01/08/202403/04/201710/12/201407/05/201407/08/2013Language:
English
Version:
01/08/2024
File size:
1.4 MB
File name:
BA00295PEN_1922-00.pdf
-
Choose from our suggested options
-
-- selected coding-- selected label99.999,00<%label%>
-
Looking for more options? Please use the individual configurator to continue
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Looking for more options? Please use the individual configurator to continue
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
- CAD drawings CAD drawings
- CAD drawings
-
Prices per piece
-
<%range%><%price%>
-
- Optional add-ons
Individual configuration
Often bought together
Accessories
Related products
We value your privacy
We use cookies to enhance your browsing experience, collect statistics to optimize site functionality, and deliver tailored advertisements or content.
By selecting "Accept all", you consent to our use of cookies.
For further details please review our cookie policy .