Title

Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.

Virus filtration (tangential flow filtration)

Explore measurement points and instrument selection across the virus filtration process

L F T P P P T A UV F A pH A Cond A pH A Cond A UV 1次側 2次側 G - ウイルスろ過 (TFF) フィルター Einlass Auslass G - Virusfiltration (Tangentialflussfiltration) Filter Inlet Outlet G - Virus filtration (tangential flow filtration) Filter
Tank level monitoring
Retentate flow
Temperature
Retentate pressure
Permeate pressure
Feed pressure
Temperature
Permeate UV
Permeate flow
System pH
System conductivity
Retentate pH
Retentate conductivity
Retentate UV
Measuring point Product Alternatives
Level measurement in: Tank level monitoring

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Cleanability, no dead legs, front flush mounting
  • No interaction with agitator operation
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: Continuous / Liquids
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Flow measurement in: Retentate flow

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Cleanability (CIP/SIP), no dead legs
  • Fully self-drainable in horizontal installations
  • Design according to ASME BPE standard
  • High reproducibility over a large flow range
  • Small quantities

Technical requirements:

  • Measuring task: Liquids
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40 °C (104 °F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Temperature measurement in: Temperature

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Cleanability, no dead legs, front flush mounting
  • Installation in thermowell
  • Easy calibration
  • Long-term stability
  • Linearity over the measurement range
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: Thermometer
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Pressure measurement in: Retentate pressure

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Front flush mounting for drainability
  • Fast temperature shock response
  • High linearity
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: Pressure
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Pressure measurement in: Permeate pressure

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Front flush mounting for drainability
  • Fast temperature shock response
  • High linearity
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: Pressure
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Pressure measurement in: Feed pressure

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Front flush mounting for drainability
  • Fast temperature shock response
  • High linearity

Technical requirements:

  • Measuring task: Pressure
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Temperature measurement in: Temperature

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Cleanability, no dead legs, front flush mounting
  • Installation in thermowell
  • Easy calibration
  • Long-term stability
  • Linearity over the measurement range
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: Thermometer
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Analysis measurement in: Permeate UV

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Fast response time
  • Cleanability (CIP/SIP), gap-free design
  • Fully self-drainable in horizontal installations
  • Design according to ASME BPE standard
  • Smallest hold-up volume

Technical requirements:

  • Measuring task: Concentration
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Flow measurement in: Permeate flow

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Cleanability (CIP/SIP), no dead legs
  • Fully self-drainable in horizontal installations
  • Design according to ASME BPE standard
  • High reproducibility over a large flow range
  • Small quantities

Technical requirements:

  • Measuring task: Liquids
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40 °C (104 °F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Analysis measurement in: System pH

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Cleanability (CIP/SIP), no dead legs
  • High accuracy
  • Suitable for CIP/SIP and autoclaves
  • Long-term stability
  • Easy and accurate calibration
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: pH
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Analysis measurement in: System conductivity

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • Cleanability (CIP/SIP), gap-free design
  • Wide measurement range 1µS/cm to 100 mS/cm
  • Suitable for CIP/SIP and autoclaves
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: Conductivity
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Analysis measurement in: Retentate pH

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • High accuracy
  • Suitable for CIP/SIP
  • Long-term stability
  • Easy and accurate calibration
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: pH
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Analysis measurement in: Retentate conductivity

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

Alternatives:

General Requirements:

  • High accuracy
  • Wide measurement range
  • Suitable for CIP/SIP
  • Fast response time
  • Design according to ASME BPE standard

Technical requirements:

  • Measuring task: Conductivity
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Analysis measurement in: Retentate UV

Endress+Hauser offers a suite of analytical instruments ideal for this stage.

General Requirements:

  • Fast response time
  • Cleanability (CIP/SIP), gap-free design
  • Fully self-drainable in horizontal installations
  • Design according to ASME BPE standard
  • Smallest hold-up volume

Technical requirements:

  • Measuring task: Concentration
  • Process pressure: 0 to 16 bar (0 to 232 psi)
  • Max. process temperature: 40°C (104°F)
  • CIP/SIP: 130°C (266°F)
  • Density: 0.8 to 1.25 g/cm³
Stainless steel process skid used in biopharmaceutical manufacturing

Key instrumentation challenges in the virus filtration process

Differential pressure as an indicator of fouling: Continuous virus filtration systems are particularly sensitive because prolonged operation increases the risk of pressure build-up and fouling. Variations in feed conditions can also increase system pressure.

Flow and flux stability as an indicator of process control: Virus filtration performance depends heavily on maintaining a defined flux through the membrane. Flow instability can affect residence time and filter loading.

Detecting filter fouling before failure: Unfortunately, there is rarely a direct measurement for the exact lifetime left in the virus filter. Fouling develops gradually and no single instrument directly measures membrane integrity during operation. This is why trend analytics are often more valuable than a single measurement.

Portfolio highlights

Explore our innovations for your virus filtration process

/
FAQ

Frequent questions on the virus filtration process