Rod probe: +/- 2 mm (0.08 in) Rope probe <= 15 m (49 ft): +/- 2 mm (0.08 in) Rope probe > 15 m (49 ft): +/- 10 mm (0.39 in) Coaxial probe: +/- 2 mm (0.08 in)
Process temperature
XT: -196...+280 °C (-321...+536 °F) HT: -196...+450 °C (-321...+842 °F)
Process pressure / max. overpressure limit
Vacuum...400 bar (Vacuum...5800 psi)
Max. measurement distance
Rod: 10 m (33 ft) Min DK>1.6 Rope: 25 m...30 m (82 ft...98 ft) Min DK>1.6; 30 m...45 m (98 ft...148 ft) Min DK >1,9 Coaxial probe: 6 m (20 ft) Min DK>1.4
Levelflex FMP54 for continuous level measurement in liquids under extreme conditions. The process connection with its ceramic-graphite seal safeguards high temperature and high pressure applications as they occur in steam boilers and toxic media like ammonia. The gas tight feedthrough guarantees additional safety. Only the gas phase compensation of the FMP54 gives reliable results in case of gas and steam phases. Reliable measurement in case of moving surface and foam or in changing medias.
Version with rod, rope or coax probe.
Process connections: Thread or flange
Temperature: -196 to +450°C (-320 to +842°F)
Pressure: -1 to +400bar (-14.5 to +5,800psi)
Maximum measuring range: Rod 10m (33ft), rope 45m (148ft), coax 6m (20ft)
Protolaunch uses Endress+Hauser flowmeters and level sensors at a rocket propulsion test facility in Buckinghamshire to demonstrate the performance of the propulsion systems and components.
Raman spectroscopy from Endress+Hauser has established itself as a key PAT tool at Evonik. It enables significant efficiency gains, cost reductions, and faster market launch of new products.
Messer produces hydrogen in a steam reformer. The facility has a large number of protection and safety devices where pressure values play an important role.
Kemira, a global chemicals company, needs accurate monitoring of level and temperature for polymer manufacturing. They use instrumentation with Bluetooth for safe and easy access to process data.
Premium device high pressure/ high temperature Rod probe, Rope probe, Coaxial Probe Integrated data memory, Factory precalibrated, Reliable measuring: in case of moved surface + foam, for changing medias
Interface measurement
Clean interfaces liquid/ liquid; Simultaneous measurement of interface and overall level
Specialities
Heartbeat Technology, Bluetooth® commissioning, Operation and maintenance SmartBlue App, HistoROM, RFID TAG for easy identification
Supply / Communication
2-wire (HART / PROFIBUS PA/ FOUNDATION Fieldbus) 4-wire (HART) Bluetooth® wireless technology and App (optional)
Accuracy
Rod probe: +/- 2 mm (0.08 in) Rope probe <= 15 m (49 ft): +/- 2 mm (0.08 in) Rope probe > 15 m (49 ft): +/- 10 mm (0.39 in) Coaxial probe: +/- 2 mm (0.08 in)
Ambient temperature
-50...+80 °C (-58...+176 °F)
Process temperature
XT: -196...+280 °C (-321...+536 °F) HT: -196...+450 °C (-321...+842 °F)
Thread: G 1 1/2, MNPT 1 1/2 Flange: ASME 2"...4", DN50...DN100, JIS 10K, 63K, Fisher Masoneilan
Sensor length
Rod probe: 10 m (33 ft) Rope probe: 45 m (148 ft) Coaxial probe: 6 m (20 ft)
Max. measurement distance
Rod: 10 m (33 ft) Min DK>1.6 Rope: 25 m...30 m (82 ft...98 ft) Min DK>1.6; 30 m...45 m (98 ft...148 ft) Min DK >1,9 Coaxial probe: 6 m (20 ft) Min DK>1.4
Communication
4...20 mA HART PROFIBUS PA FOUNDATION Fieldbus Bluetooth® wireless technology
Certificates / Approvals
ATEX, FM, CSA C/US, IEC Ex, INMETRO, NEPSI, KC, EAC, JPN Ex, UK Ex
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex